EDX-9000MAX Energy-Dispersive X-ray Fluorescence Spectrometer
Advantages of the Instrument The EDX-9000MAX Energy-Dispersive X-ray Fluorescence Spectrometer offers excellent stability, fast and accurate analysis, low operating costs, and easy operation and maintenance. It can detect all elements from Na(11) to U(92). Its main advantages are as follows: ...
The EDX-9000MAX Energy-Dispersive X-ray Fluorescence Spectrometer offers excellent stability, fast and accurate analysis, low operating costs, and easy operation and maintenance. It can detect all elements from Na(11) to U(92).
Its main advantages are as follows:
- Utilizes fully digital spectral analysis technology (a method, system, and digital spectrometer for high-speed digital separation of nuclear pulse signals), featuring high counting rates, zero missed counts, and excellent stability
- Equipped with state-of-the-art electrically cooled silicon drift semiconductor detectors, enabling the analysis of a wide range of elements with high resolution
- Features a high-vacuum sample measurement chamber, which eliminates the attenuation of low-energy X-rays by air, ensuring accurate analysis of light elements
- Sample types: powder, bulk
- Chamber environment: air or vacuum
- Features automatic peak detection and other methods to ensure both short-term and long-term instrument stability
- Automatically switches filters and collimators based on sample type to ensure measurement accuracy for different samples
- Open-platform calibration system for calibration curves, allowing for customized substance detection, analysis, and control solutions
- Customizable report formats (Excel, PDF, etc.) to meet various statistical and formatting requirements
- User-friendly, integrated design adapts to various environments, supports 24/7 continuous operation, and features a low failure rate
- 2048-channel multichannel spectrometer provides real-time measurement and display, enabling detailed analysis of sample X-ray spectra; the system automatically identifies spectral lines, facilitating understanding of sample composition
- Enables rapid, non-destructive, and accurate simultaneous analysis of multiple elements
Data management includes user passwords to enhance the integrity of test data and the reliability of quality control.
| Parameter Category | Parameter Item | Parameter Specification |
|---|---|---|
| Detector | FASTSDD electro-cooled detector | Imported electro-cooled detector |
| Amplifier circuit module | Detects characteristic X-rays from the sample and further amplifies the acquired data | |
| Detector window | Be window | |
| Detector area | 25mm², large-area window that increases the signal reception area | |
| Energy resolution | 129±5eV | |
| Peak-to-background ratio | >20,000:1 (count ratio at 5.9keV vs. 1keV) | |
| X-ray Generation System (Thin Beryllium Window X-ray Tube) | Cathode filament | Maximum current limit:1.8A |
| Target Material | Standard vacuum-sealed Rh target;for high-precision testing of Na and Cl elements,an Ag target is recommended | |
| Voltage Range | 5–50kV | |
| Current Range | 0–1mA | |
| Maximum Power | 50W | |
| Stability | Less than 0.02% per 8 hours after 30 minutes of operation | |
| Service life | ≥10,000hours | |
| High-Voltage Power Supply | Voltage range | 5–50kV |
| Current range | 0–2mA | |
| Maximum power | 50W | |
| Stability | Less than 0.02% per 8 hours after 30 minutes of operation | |
| Ripple Voltage | At rated output voltage,the peak ripple voltage is 0.1% of the maximum voltage | |
| Dimensions and Weight | Overall Dimensions | Length:80cm, Width:100cm, Height:160cm |
| Weight | ≤150kg (main unit) | |
| Operating Environment | Operating Temperature | 10–30°C (air conditioning recommended) |
| Operating Humidity | (30–70)% R.H. (non-condensing) | |
| Power Supply | Single-phase 200–240V @ 50Hz,current>5A,single-phase three-wire power circuit (main unit;other accessories calculated separately) | |
| Instrument Parameters | Element Testing Range | Na(11)–U(92) |
| Element concentration range | 2ppm to 99.99% | |
| Test duration | Selectable from 30 to 300 seconds | |
| Instrument detection limit (Fe) | As low as 2ppm | |
| Instrument measurement accuracy | Repeatability 0.05%, Stability 0.05% (at 96% Fe concentration) | |
| Note | Remark | For reference only; actual configuration will be based on customer requirements. |
Smelting raw materials: iron ore, magnetite concentrate, manganese ore, zinc concentrate, lead concentrate, bauxite, remelt slag, nickel catalyst, etc.
Please use our online inquiry contact form below if you have any questions, our team will get back to you as soon as possible.